This series of steam sterilizable electrodes is designed for economical
use. They are constructed very similarly to the standard pH electrodes,
have a 10 and l2mm body, and can be ordered with the desired "H" length.
"H" length is determined by the insertion depth on the under side of the
fermentor headplate to the tip of the probe. Glass disposable electrodes
are available in both polarographic and galvanic styles to fill all requirements.
The glass DO probes are available for laboratory fermentation, where
sterilization takes place in a separate autoclave. The disposable pHoenix
Dissolved Oxygen Electrodes are compatible with most common oxygen amplifiers
found in fermentation. Glass body oxygen electrodes are non-temperature
compensated and are required to be kept at a constant temperature. Calibration
must be made at the same temperature as the controlled fermentation. The
electrodes feature external filling tubes. Electrolyte can be added during
use without removing the electrode from the vessel. The electrode membranes
can be replaced very easily. However, some skill is involved in replacing
them.
The polarographic type electrodes are to be used with Ingold, pHoenix
amplifiers or amplifiers from other manufacturers designed to be used with
Ingold DO electrodes. The galvanic type electrodes are designed to be used
with NBS, LSL Biolafitte and pHoenix Electrode Co. amplifiers.
| TYPE |
027IP15 |
027BG15 |
027NG15 |
| CATHODE |
Pt |
Au |
Au |
| ANODE |
Ag/AgCl |
Pb |
Pb |
| RESPONSE |
98% 60 seconds |
98% 90 seconds |
98% 90 seconds |
| STABILITY* |
Better than 2% |
Better than 2% |
Better than 2% |
| TEMPERATURE COMPENSATION |
No |
No |
No |
TEMPERATURE
RANGE |
0~60°C |
0~60°C |
0~60°C |
| STERILIZATION TEMPERATURE |
0~130°C |
0~130°C |
0~130°C |
| RESIDUAL SIGNAL |
Less than 1nA |
0.5 ~ 1mV |
Less than 0.4µA |
| OUTPUT** |
50nA |
40mV ± 5mV |
10 ~ 15µA |
*Constant temperature.
**Medium saturated with air.
ORDERING INFORMATION
| PART NO. |
"H" LENGTH |
BODY DIAMETER |
TYPE |
AMPLIFIER COMPATIBILITY |
| 027IP15 |
"H"-DL |
12mm |
Polarographic |
pHoenix, Ingold & others |
| 027BG15* |
"H"-DL |
12mm |
Galvanic |
LSL Biolafitte |
| 027NG15 |
"H"-DL |
10mm |
Galvanic |
New Brunswick Scientific |
*Must be used with 9001094-010U cable
MEMBRANE SERVICE KITS AND FILLING SOLUTIONS
| M0500T1-K5 |
Membrane service kit K-5 of both glass polarographic and galvanic electrodes.
Consists of 10 membranes 1 mil thick, 10 silicone bands. |
| M0500T2-K5 |
Membrane service kit K-5 of both glass polarographic and galvanic electrodes.
Consists of 10 membranes 2 mil thick, 10 silicone bands. |
| R001068 |
Polarographic electrode filling solution, 125ml. |
| R001069 |
Galvanic electrode filling solution, 125ml. |
C1 TYPE CABLES - Standard 3m lead for non-temperature compensated electrodes.
| 9001091-010B |
BNC connector |
| 9001091-010I |
DIN connector for use with Ingold pO2 amplifier |
| 9001091-010D |
Dual pin connector for use with NBS glavanic pO2 amplifier |
| 9001091-010U |
Banana plugs for use with with LSL Biolafitte pO2 amplifier |
| 9001091-010T |
No connector for Phoenix, Applikon & Cole Parmer pO2
amplifiers or for user supplied connector. |
| 9001091-010CC |
Custom cable with choice of plug, lemo connector. Specify size and
pin number. |
| 9001093-010D |
Pin connector for use with 025NG15 & 027NG15 galvanic O2
electrode. Cable features a load resistor for direct connection to a recorder
with mV input. (Approx. 40mV staurated air/water). |
|