This
series of steam-sterilizable electrodes is designed for small fermentors
where sterilization is performed in an autoclave in non pressurized operation.
The electrodes can withstand repeated sterilization up to 130°C and
feature a detachable cable assembly insulated with a water-tight connector.
The electrodes are autoclaved with the cable detached and a protective
cap over the connector. Autoclavable electrodes, with detachable leads,
are generally used with Ingold probe assemblies 761-351, 761-31, 761-361,
761CH-31, and Radiometer electrode mountings commonly used in fermentation
processes. Electrode construction features an annular ceramic liquid junction
to eliminate junction problems which normally occur with autoclavable steam-sterilizable
electrodes. Electrodes are available in both single and double junction
reference reservoir. The double junction reference style is recommended
in most cases for long-term stability and life expectancy. Both reference
reservoirs are refillable with a special high-temperature reference electrolyte.
When ordering, the detachable cable must be ordered seperately with the
proper pH meter connection. Unless otherwise specified, the immersion length
is 150 mm. The immersion length "A" is determined by the immersion depth
"H" of the probe plus 70 mm. (Except Ingold probe 761-31, A=H+50 mm).
FEATURES
| 1 |
Removable electrode cable connection. |
| 2 |
Internal Ag/AgCl reference chamber fill port "A".
(On double junction reference reservoir only.) |
| 3 |
Electrolyte bridge fill Port "B". |
| 4 |
Internal annular ceramic reference junction.
(On double junction reference reservoir only.) |
| 5 |
Annular ceramic primary reference junction |
| 6 |
pH bulb or platinum band sensor |
| 7 |
12 mm O.D. shaft. |
SPECIFICATIONS
| Part Number |
Description |
Sensor
/Glass |
Sensor
Range |
Temperature
Range |
| A773593-???-DL |
pH Single reference reservoir |
GX-2 |
0 - 13pH |
-5 to 135°C |
| A774519-???-DL |
pH Single reference reservoir |
GX-3 |
0 - 14pH |
40 to 135°C |
| A773792-???-DL |
pH Double reference reservoir |
GX-2 |
0 - 13pH |
-5 to 135°C |
| A774713-???-DL |
pH Double reference reservoir |
GX-3 |
0 - 14pH |
40 to 135°C |
| A779685-???-DL |
ORP Single reference reservoir |
Platinum |
±1000mV |
-5 to 135°C |
| A779861-???-DL |
ORP Double reference reservoir |
Platinum |
±1000mV |
-5 to 135°C |
Replace ??? in part number with length "A" in mm.
An adapter #8001061 is required to convert the DL connector to Ingold